DP Flow Transmitter Calibration: A Step-by-Step Guide

 14 June 2023

Differential pressure flow meters are widely used in the industry for the measurement of flow in a closed pipe.

This kind of flowmeter consists of two parts:

The DP flow meter will calculate the flow rate by measuring the differential pressure across a primary element.

Various types of primary elements can be used to obtain differential pressure. Some of the most common primary elements are orifice plates, (averaging) pitot tubes, venturi tubes, and flow nozzles.

They all operate on the same principle that the flow through the primary element is proportional to the differential pressure across the element.

Although DP type flow transmitters are known for their high performance and reliability, they will need a recalibration from time to time.

For the purpose of the calibration procedure that we will describe further on, we will consider a DP transmitter with an orifice plate as the primary element.

Orifice DP Flow Transmitter

To be clear, we are not going to calibrate the entire flow measurement, but only the DP flow transmitter. To calibrate the complete flow measurement, the primary element and the DP flow transmitter would have to be demounted and then installed on a flow test bench to be compared with a reference flow meter. This is something that is typically done in a calibration lab.

Relationship between Flow Rate and Differential Pressure

The following equation describes the relationship between flow rate and differential pressure:

Formula for the relationship between flow rate and differential pressure.

where Q is the volumetric flow rate, k is a constant, DP is the differential pressure, and the square root function represents the relationship between velocity and pressure.

From this equation, we can see that an increase in differential pressure will result in an increase in flow rate. However, the relationship between flow rate and differential pressure is not linear but square-rooted, as you can see from the diagram below.

table DP vs Flow rate
DP vs Flow rate curve

A 25% increase in differential pressure will result in a 50% increase in flow rate.

25% DP corresponds to 50% flow rate

Low-flow Cut-off

Let’s have another look at the DP vs Flow curve above. Check in particular how the curve behaves in the region around the origin.

In this region, the slightest change in differential pressure causes the flow to change rapidly. A little noise on the DP signal gets multiplied by a factor of 10 or more on the flow signal.

This makes the flow signal very unstable and unreliable. Therefore, vendors will define a low-flow cut-off for the flow meter. That means that the output signal will be set to 4mA once the flow drops below 10%.

The low-flow cut-off is only present in DP transmitters with internal square root extraction. If the square rooting is done by the PLC/DCS, then the DP transmitter has a linear output and the low-flow cut-off should be programmed in the PLC/DCS.

A low-flow cut-off at a flow rate of 10% corresponds to an output of 5.6 mA. So, between 4 and 5.6 mA there is no useful measurement possible because of too much inaccuracy.

All of that will have consequences on the calibration of our DP flow transmitter. For a typical five-point calibration of a transmitter with linear output, the test points would be 0, 25, 50, 75, and 100% of the flow rate. But, if the DP flow transmitter performs the square root calculation and a Low-Flow Cut-Off is considered, it has no sense to have a test point below 10% of the flow rate. So, the test points should be 10, 25, 50, 75, and 100% of the flow rate.

table DP vs Flow rate with low-flow cut-off
DP vs Flow rate curve with low-flow cut-off

What are the tools needed to calibrate a DP type flow transmitter?

Besides the DP flow transmitter itself, we will need the below-mentioned tools to perform a calibration:

Some important things to consider before starting the calibration of a DP-type flow transmitter

The DP type flow transmitter calibration procedure

The setup for a DP flow transmitter calibration

When all the preparatory work has been completed, we can now start calibrating the DP flow transmitter.

  1. Arrange all the tools as shown in the above figure.
  2. The DP-type power transmitter can be powered through the field cables connecting the DP transmitter to the system, or by using an external power supply shown in the figure.
  3. If the differential pressure flow transmitter is in the field, then isolate the DP transmitter. Flush the differential pressure transmitter in a closed drain if it comes into contact with a dangerous process medium like a hazardous chemical.
  4. Connect the power supply, the multimeter, and the DP flow transmitter in a series configuration. Make sure to connect the multimeter leads to the mA terminal and the common terminal.
  5. Now connect the DP transmitter to the pressure calibrator using the appropriate tubing. Here we can use PU tubing because the pressure to be applied to the DP flow transmitter is in the mmH2O range.
  6. Connect a precision pressure gauge of appropriate range to the pressure-generating device if for any reason this device would not display the generated pressure.
  7. Prior to starting the calibration, it is necessary to determine whether the square root extraction takes place in the DP transmitter or in the PLC/DCS. The output value (mA) of the DP transmitter will depend on where this square root extraction occurs. So, check the datasheet or the configuration parameters of the DP transmitter to see if the transmitter does the square root extraction. A third option is to simulate 25% of the DP range with the pressure calibrator and look at the output (mA) of the DP transmitter:
    • If the output is about 12mA, the sqrt extraction is performed by the transmitter.
    • If the output is about 8mA, the sqrt extraction is performed by the PLC/DCS.
  8. Let’s take an example to see what we can expect. Imagine a DP flow meter with the following ranges:
    * Flow range: 0 – 1000 m3/h
    * DP range: 0 – 5000 mmH2O
    We want to do a five-point calibration of the transmitter‘s mA output.
    Now there are 2 possibilities: the square root is done in the transmitter, or it’s done in the PLC/DCS.
    Let’s have a look at both possibilities.
    • If the transmitter performs the square root extraction, we have a mA output representing the flow rate.
      So since we want to calibrate the transmitter’s output, we will have to distribute our test points evenly over the output range. Considering what we said earlier about the inaccuracy at the low flow range, we will test for 10%, 25%, 50%, 75%, and 100% of the flow range.
      We will arrange these values in the table below and add the corresponding DP-values and mA-signals.
Test pointsFlow (%)Flow (m3/h)DP (mmH2O)Output (mA)
TP110 %100505,6
TP225 %250312.58
TP350 %500125012
TP475 %7502812.516
TP5100 %1000500020

The values in the DP (mmH2O) and Output (mA) columns have been calculated with the following formulas:

Formulas to calculate DP and mA

Flow value = flow rate at the test point

Flow span = 1000

DP span = 5000

Test pointsFlow (%)DP (mmH2O)Flow (m3/h)Output (mA)
TP10 %004
TP225 %12505008
TP350 %2500707,112
TP475 %375086616
TP5100 %5000100020

The values in the Flow (m3/h) and Output (mA) columns have been calculated with the following formulas:

Formulas to calculate Flow and mA

DP value = DP at the test point

DP span = 5000

Flow span = 1000

  1. Now that we know where the square root extraction is done and we have calculated our test points, we can start the validation procedure. We will do an ‘As Found’ test on all test points.
    This test is done for two reasons:
    • If this test proves that the instrument is still within the required accuracy for the application then we do not have to recalibrate it.
    • With this test, we will record the history of the device performance data. This will be helpful to decide if we need to shorten or prolong the calibration frequency.
Test
Points
AppliedTransmitter outputError
%reading
DP
(mmH2O)
Flow
(m3/h)
DP
(mmH2O)
Flow
(m3/h)
50
312,5
1250
2812,5
5000
2812,5
1250
312,5
50
  1. Start the validation procedure with zero pressure on the pressure calibrator and increase the pressure until you reach the pressure value of the first test point. Note down the reading from the pressure calibrator in the second column ‘Applied DP’ and the reading from the transmitter in the fifth column ‘Transmitter output Flow’.
    Similarly, apply the pressures of the other test points in the order they appear in the table. Note down the readings from the pressure calibrator and the transmitter in the appropriate columns.
    Every time you reach a new test point, the applied pressure value should remain stable for 30 seconds before reading the output value from the transmitter. If the pressure drops during these 30 seconds, then there is a leakage in the tubing or an issue with the pressure calibrating device. Identify the root cause and attend to it.
  2. As the square root extraction is done by the transmitter, its output represents the flow rate. To calculate the error on the output, we will have to compare the measured flow rate with the applied flow rate. As we can only apply a DP, we will have to calculate the applied flow with the following formula:
Formula to calculate flow
  1. Now calculate the error % from the Flow values we got with the next formula:
formula to calculate error percentage
  1. If the calculated errors are all well within the required accuracy for the application, then the DP type flow transmitter is ok and should not be recalibrated. If the error obtained from the calculations is more than the required accuracy for the application then a calibration of the DP flow transmitter is needed.
  2. A recalibration of a dp flow transmitter with internal square root extraction consists of a lower trim at the 10% flow point and a span trim at the 100% flow point. We can not do a zero trim for a transmitter with square root extraction because the zero point is too unstable. Any fluctuation in the input signal, no matter how small, is multiplied tenfold at the output.
    First, apply 1% DP on the input of the transmitter and check if this corresponds to 10% flow at the output. If this is not ok then start a lower trim at the 1% input / 10% output level (not zero) using the HART communicator.
  3. Now check the 100% flow point by applying the 100% DP pressure to the input of the transmitter. If the reading does not indicate 100% flow, perform a span trim using the HART calibrator.
  4. When the lower trim and span trim is completed, the transmitter is calibrated. To check whether everything went well and whether the transmitter will measure correctly, we will now do an ‘As Left’ verification.
    Repeat the procedure from step 9 to the end. Set up the ‘As Left’ table with the same test points as those from the ‘As Found’ table.
    If the calculated errors from the ‘As Left’ table fall within the required accuracy for the application, the calibration is successful.
  5. All that remains is to return the transmitter to service. Inform the operator that the DP flow transmitter calibration job is finished and remove any interlock bypasses you placed at the start.

Why a differential pressure flow transmitter needs to be calibrated

All instruments need to be recalibrated over time. A DP flow meter makes no exception to that rule. A slight progressive accuracy degradation also called drift is the reason for these recalibrations. Drift is caused by the aging of the material and the wear and tear on mechanical and electronic components.

The sensor of the transmitter can be considered as a mechanical component. During its lifetime, this sensor will be subjected to constant pressure fluctuations, causing it to be repeatedly tensioned and released. This can eventually lead to small deformations causing small inaccuracies. The same can happen to the diaphragm that separates the sensor from the process, the body of the pressure transmitter, …

The transmitter’s electronics also age under the influence of temperature fluctuations, humidity, mechanical shock, or vibration… which changes the processing of signals over time.

To compensate for all these small changes, we occasionally need to recalibrate the transmitter.

Factors affecting the DP flow transmitter calibration frequency

How often should you perform a DP flow transmitter calibration?

When you install a DP flow meter on a new application, you want to have an idea of when you will need to recalibrate it. Well, there is a way to determine the calibration interval by calculation.

I’ll show you how to do it.

First, you need to decide on the accuracy required for your application, and by required accuracy I don’t mean the reference accuracy of the DP transmitter but the total performance of the installed measurement.

Errors that occur through ambient temperature changes and high static pressure on the differential pressure transmitter should therefore be taken into account.

Generally, for pressure transmitters, the total installed performance will be between 0,5 and 2,0% of calibrated span. Here is a short list of applications with their required installed performance to help you make your choice:

● Plant safety and efficiency control0,5%
● Regulatory control1,0%
● Supervisory control1,5%
● Monitoring2,0%

Once you have determined the installed performance, I want you to calculate the total probable error with this formula:

formula to calculate the total probable error

To populate the equation you will need to dive into the published specifications of the differential pressure transmitter.

The reference accuracy will be given in % of span. Find that number and put it in the formula.

The ambient temperature effect can be given with a formula where a part of it is expressed in the unit ‘% URL’ and another part in the unit ‘% span’. When this is the case, you should convert the unit ‘% URL’ to the unit ‘% span’ as follows:

formula to convert the %URL to %span

The static pressure effect also called the line pressure effect, could be given as a % of reading. Because we have to convert everything to % of span, we will simply consider the pressure reading at full span. This way, we can just keep the percentage value and replace the unit ‘% of reading’ with ‘% of span’.

Now, populate the above formula with the 3 results we have found and calculate the Total Probable Error.

The next thing to do is to search for the stability specification of the output. This will be given as a % of URL for a number of years, meaning that the instrument will lose accuracy at a certain % of URL over the stated number of years.

You should now convert the stability specification to engineering units, like e.g. inH2O, by replacing URL with its pressure value. This gives the stability in engineering units over a number of years. From this, you can calculate the stability per month instead of over a number of years.

Finally, you have to convert the required accuracy for your application and the total probable error to engineering units by replacing the word span with the value of the span.

By using the following formula with the engineering units, you will find the calibration interval in months.

formula to calculate the calibration interval
Advertisement

2 Comments

  1. Mboa Djidji Arnaud says:

    I appreciate your lessons. I want to learn instrumentation

Leave a Comment

Your email address will not be published. Fields marked with * are required.

bookmark invitation